EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 222

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
205Operating Manual 1407.6062.32 ─ 08
"Ext Triggered"
The pulse generator signal is triggered by an external trigger event. The
trigger signal is supplied via the PULSE EXT connector.
Example: Generation of pulse signals using trigger modes Ext
Triggered (Single Pulse)
The measurement were made using a 6-dB-attenuator.
Parameter Value
Trigger Mode Ext Triggered
Double Pulse State Off
External Trigger Input Slope Positive
Pulse Delay 100 ns
Pulse Width 100 ns
Pulse Period -
Channel
1
= Indicates the external trigger signal which is input at the PULSE EXT con-
nector. The positive slope is active.
Channel
2
= Indicates the sync signal. The sync signal starts after a trigger delay of typi-
cally 50 ns (see specifications). The sync signal is output at the PULSE
SYNC connector.
Channel
3
= Indicates the pulse signal. The pulse signal starts after the set pulse delay of
100 ns. It is output at the PULSE VIDEO connector.
Example: Generation of pulse signals using trigger modes Ext
Triggered (Double Pulse)
The measurement were made using a 6-dB-attenuator.
Parameter Value
Trigger Mode Ext Triggered
Double Pulse State On
External Trigger Input Slope Negative
Pulse Width 100 ns
RF Signal and Analog Modulations - RF Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A