Instrument Function
R&S
®
SMBV100A
220Operating Manual 1407.6062.32 ─ 08
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:EXEC
SCPI command:
[:SOURce<hw>]:SWEep:POWer:EXECute on page 699
Use Digital Attenuation
Activates the use of digital attenuation during RF power sweep.
SCPI command:
[:SOURce<hw>]:SWEep:POWer:DATTenuator:STATe on page 698
Start Level - Level Sweep
Sets the start level.
SCPI command:
[:SOURce<hw>]:POWer:STARt on page 683
Stop Level - Level Sweep
Sets the stop level.
SCPI command:
[:SOURce<hw>]:POWer:STOP on page 684
Current Level - Level Sweep
Displays the current level.
If "Step" is set, the level for the next level step of the sweep is entered here.
SCPI command:
[:SOURce<hw>]:POWer:MANual on page 679
Shape - RF Level Sweep
Selects the cycle mode for a sweep sequence (shape).
"Sawtooth"
One sweep runs from the start level to the stop level. The subsequent
sweep starts at the start level again, i.e. the shape of sweep sequence
resembles a sawtooth.
"Triangle"
One sweep runs from start to stop level and back, i.e. the shape of the
sweep resembles a triangle. Each subsequent sweep starts at the start
level again.
SCPI command:
[:SOURce<hw>]:SWEep:POWer:SHAPe on page 700
Step - Level Sweep
Sets the step width for the individual sweep steps. This entry is effective for all sweep
modes.