Remote Control Commands
R&S
®
SMBV100A
679Operating Manual 1407.6062.32 ─ 08
Parameters:
<LevMode> NORMal | LOWNoise | LOWDistortion
NORMal
The RF signal is output in the standard values of the instrument.
LOWNoise
A very low noise sinewave signal is output.
LOWDistortion
A very pure sinewave signal is output.
Example:
POW:LMODE LOWD
sets the LOWDistortion mode. The instrument reduces distortions
of the RF signal to a minimum.
[:SOURce<hw>]:POWer:MANual <Manual>
In Sweep mode (:SOUR:POW:MODE SWE) the command sets the level for the next
sweep step in the Step sweep mode (:SOUR:SWE:POW:MODE MAN). Here only level
values between the settings [:SOUR]:POW:STAR and [:SOUR]:POW:STOP are per-
mitted. Each sweep step is triggered by a separate :SOUR:POW:MAN command.
As with the "Level" value entered in the "RF Level" menu, the OFFSet value is also taken
into consideration with this command.
The specified value range is therefore only effective if :SOURce:POWer:OFFSet is set
to 0. The value range for other OFFset values can be calculated using the following
formula:
Minimum level + OFFSet ... Maximum level + OFFSet
Parameters:
<Manual> float
Minimum level ... Maximum level
The value range for the level setting varies according to the instru-
ment model
The values are given in the data sheet.
Increment: 0.01 dBm
*RST: -30 dBm
Default unit: dBm
Example:
POW:SWE:MODE MAN
sets the Step sweep mode for RF output A.
POW:MAN -5 dBm
sets an RF level of -5 dBm for the next setting in the Step sweep
mode for RF output A.
POW:MODE SWE
sets the Level Sweep mode for RF output A.
POW:MAN -5.5 dBm
triggers the next sweep step with a level of -5.5 dBm.
Usage: SCPI confirmed