EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 695

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100A
678Operating Manual 1407.6062.32 ─ 08
Parameters:
<Rcl> INCLude | EXCLude
INCLude
The stored level is also loaded when a stored instrument config-
uration is loaded.
EXCLude
The RF level is not loaded when a stored instrument configuration
is loaded. The current level is retained.
*RST: no default
Example:
POW:RCL INCL
the stored level is set if the Recall command is called.
Usage: SCPI confirmed
[:
SOURce<hw>]:POWer:LIMit[:AMPLitude] <Amplitude>
The command limits the maximum RF output level in CW and SWEEP mode. It does not
influence the "Level" display or the response to the POW? query command.
The value is not affected by an instrument preset (PRESET key), *RST and the Save/
Recall function. This parameter is influenced only by the factory preset (SYST:FPR) and
its factory value is equal to the upper limit.
Parameters:
<Amplitude> float
Minimum level ... Maximum level
The value range for the level setting varies according to the instru-
ment model.
The values are given in the data sheet.
Increment: 0.01 dB
*RST: 30 dBm
Default unit: dBm
Example:
SOURce:POWer:LIMit:AMPLitude 10
or
:POW:LIM 10
limits the RF level to maximum +10 dBm.
Usage: SCPI confirmed
[:SOURce<hw>]:POWer:LMODe <LevMode>
The command sets the RF level mode.
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals