EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 273

Rohde & Schwarz SMBV100A
780 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
256Operating Manual 1407.6062.32 ─ 08
Fig. 5-16: Effect of a negative quadrature offset
SCPI command:
[:SOURce<hw>]:BB:IMPairment:QUADrature[:ANGLe] on page 588
I/Q Skew - Digital Impairments
Sets the time offset between the I and Q vectors.
In an I/Q modulator without I/Q skew, the I and Q vectors are aligned to the marker. With
an I/Q skew, both vectors are shifted relative to the marker so that the offset of each of
the vectors to the marker will be the half of the I/Q skew value.
A positive I/Q skew means that the Q vector delays relative to the I vector and vice versa.
SCPI command:
[:SOURce<hw>]:BB:IMPairment:SKEW on page 589
IQ Delay
Sets the time delay of both I and Q vectors relative to the selected trigger and marker or
relative to the other instrument(s) working in the master-slave synchronous mode (see
chapter 5.8.2.4, "Synchronous Signal Generation", on page 277).
A positive value means that the I and Q vectors delay relative to the marker/trigger or to
the other instrument and vice versa.
SCPI command:
[:SOURce<hw>]:BB:IMPairment:DELay on page 586
Impairment of Digital I/Q Signal and Noise Generator - AWGN/IMP Block