Instrument Function
R&S
®
SMBV100A
319Operating Manual 1407.6062.32 ─ 08
In the "Retrigger" mode every trigger signal causes signal generation to restart. This
restart is inhibited for the specified number of samples.
SCPI command:
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:INHibit on page 559
Marker Mode
The marker output signal for synchronizing external instruments is configured in the
Marker Settings section "Marker Mode".
Marker x - Digital Modulation
Selects a marker signal for the associated MARKER output.
"CList"
A marker signal that is defined in the selected control list is generated.
"Pulse"
A regular marker signal is generated. The clock frequency is defined by
entering a divider. The frequency is derived by dividing the sample rate
by the divider. The input box for the divider opens when "Pulse "is
selected, and the resulting pulse frequency is displayed below it.
SCPI command:
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:DIVider on page 562
"Pattern"
A marker signal that is defined by a bit "pattern" is generated. The
"pattern" has a maximum length of 64 bits and is defined in an input
field which opens when "pattern" is selected.
SCPI command:
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PATTern on page 562