Instrument Function
R&S
®
SMBV100A
374Operating Manual 1407.6062.32 ─ 08
"Target Crest"
Optimization of the crest factor to a desired value for all carrier config-
urations. The optimization time depends on the number of carriers and
the desired crest factor. Computation time increases only when the
number of carriers exceeds 256 and the crest factor is below 4 dB. The
desired value can be entered in "Desired Crest Factor".
Note: Optimization can be cancelled at any time, and the current value
being displayed at that moment is then used.
SCPI command:
[:SOURce<hw>]:BB:MCCW:CFACtor:MODE on page 595
Desired Crest Factor
Enters the desired crest factor.
This is only possible when the optimization "Target Crest" has been selected.
SCPI command:
[:SOURce<hw>]:BB:MCCW:CFACtor on page 594
Table Setup Assistant
The "Table Setup Assistant" section can be used to set a selectable carrier range. The
carrier table can be edited in the "Carrier Table" dialog.
Carrier Start
Defines the start index of the carrier range to which the following settings are intended
to apply.
SCPI command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STARt on page 598
Carrier Stop
Defines the stop index of the carrier range to which the following settings are intended to
apply.
SCPI command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STOP on page 598
Carrier State
Switches the carriers in the carrier range on/off.
SCPI command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STATe on page 598
Power Start
Sets the power of the starting carrier.
SCPI command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer[:STARt] on page 597
Power Step
Sets the width of the step with which the power will be changed from carrier to carrier.
Baseband Signal - Baseband Block