Remote Control Commands
R&S
®
SMBV100A
483Operating Manual 1407.6062.32 ─ 08
:OUTPut<hw>[:STATe]:PON...........................................................................................485
:OUTPut<hw>:AFIXed:RANGe:LOWer?
The command queries the minimum level which can be set without the attenuator being
adjusted (Attenuator FIXed).
Return values:
<Lower> float
Default unit: dBm
Example:
OUTP:AFIX:RANG:LOW
queries the minimum level for the FIXed setting.
Example:
Response: -50
The minimum level is -50 dBm.
Usage: Query only
:OUTPut<hw>:AFIXed:RANGe:UPPer?
The command queries the maximum level which can be set without the attenuator being
adjusted (Attenuator FIXed).
Return values:
<Upper> float
Default unit: dBm
Example:
OUTP:AFIX:RANG:UPP
queries the maximum level for the FIXed setting for the RF output.
Example:
Response: -27
The maximum level is -27 dBm.
Usage: Query only
:OUTPut<hw>:AMODe <Amode>
The command switches the mode of the attenuator at the RF output (Attenuator MODe).
Parameters:
<Amode> AUTO | FIXed
AUTO
The electronically switching attenuator switches with a 5 dB step
width at fixed switching points.
FIXed
The attenuator is fixed at the current position. The uninterrupted
level settings are made if automatic level control is activated
(SOURce:POWer:ALC ON).
*RST: AUTO
OUTPut Subsystem