Remote Control Commands
R&S
®
SMBV100A
484Operating Manual 1407.6062.32 ─ 08
Example:
POW:ALC ON
activates automatic level control for RF output.
OUTP:AMOD FIX
sets the fixed mode with uninterrupted level for RF output.
:OUTPut<hw>:IMPedance?
The command queries the impedance of the RF outputs. This permits converting the
output level between units V and W. The impedances cannot be changed.
Return values:
<Impedance> G50 | G1K | G10K
Default unit: Ohm
Example:
OUTP:IMP
queries the impedance of RF output.
Response: 50
the impedance is 50 ohms
Usage: Query only
SCPI confirmed
:OUTPut<hw>:PROTection:CLEar
The command resets the protective circuit after it has been tripped. The state of the output
is again determined by OUTPut:STATe.
Example:
OUTP:PROT:CLE
resets the protective circuit for RF output.
Usage: Event
SCPI confirmed
:OUTPut<hw>:PROTection:TRIPped?
The command queries the state of the protective circuit.
Return values:
<Tripped> 0 | 1 | OFF | ON
Example:
OUTP:PROT:TRIP
Queries the state of the protective circuit for RF output A.
Response: 0
The protective circuit has not tripped.
Response: 1
The protective circuit has tripped.
Usage: Query only
:OUTPut<hw>[:STATe] <State>
This command activates and deactivates the RF output.
OUTPut Subsystem