Remote Control Commands
R&S
®
SMBV100A
488Operating Manual 1407.6062.32 ─ 08
Example:
SENS1:DISP:PERM:STAT ON
the permanent viewer is switched on.
SENS1:DISP:PERM:PRI AVER
the measured average power is indicated.
SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
The command queries the current filter length for auto filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE AUTO)
Return values:
<Auto> float
Example:
SENS1:FILT:TYPE AUTO
selects auto filter mode for the power sensor connected to the
SENSOR connector.
SENS1:FILT:LENG:AUTO?
queries the automatically set filter length.
Response: 1024
Usage: Query only
SENSe<ch>[:POWer]:FILTer:LENGth[:USER] <User>
The command selects the filter length for user filter mode
(SENSe:POWer:FILTer:TYPE USER). As the filter length works as a multiplier for the
time window, a constant filter length results in a constant measurement time. Values 1
and 2^n are settable.
The time window is fixed to 20 ms.
Parameters:
<User> float
Range: 1 to 65536
*RST: 1
Example:
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:LENG 16
sets a filter length of 16. The resulting measurement time is 640
ms (2x16x20 ms).
SENSe<ch>[:POWer]:FILTer:NSRatio <Nsratio>
The command defines the noise content for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value determines the
proportion of intrinsic noise in the measured result.
SENSe, READ and INITiate Subsystems