Remote Control Commands
R&S
®
SMBV100A
489Operating Manual 1407.6062.32 ─ 08
Parameters:
<Nsratio> float
Range: 0.0001 to 1.0
*RST: 0.01
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR 0.2
sets a noise content of 0.2.
SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe <Mtime>
The command defines the timeout for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value ensures limited
settling times.
Parameters:
<Mtime> float
Range: 1.0 to 999.99
*RST: 4
Default unit: s
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR .2
sets a noise content of 0.2.
SENS1:FILT:NSR:MTIM 5
limits the settling time to 5 seconds
SENSe<ch>[:POWer]:FILTer:SONCe
The command activates the search for the optimum filter length for the current measure-
ment conditions. The found filter length can be retrieved with com-
mand :SENSe:POWer:FILTer:LENGth:USER?. This command is only available for
user filter mode (:SENSe:POWer:FILTer:TYPE USER).
Example:
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:SONC
activates the search for the optimum filter length.
SENS:FILT:LENG?
returns the found optimum filter length.
Response: 128
Usage: Event
SENSe, READ and INITiate Subsystems