Remote Control Commands
R&S
®
SMBV100A
509Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:BB:ARB:TRIGger[:EXTernal<1|2>]:INHibit................................................514
[:SOURce<hw>]:BB:ARB[:TRIGger]:SEQuence................................................................514
[:SOURce<hw>]:BB:ARB:TRIGger:ARM:EXECute
The command stops waveform output for trigger modes Armed_Auto and Armed_Ret-
rigger. A subsequent internal or external trigger event restart waveform output.
Example:
BB:ARB:TRIG:SOUR INT
sets internal triggering.
BB:ARB:TRIG:SEQ ARET
sets Armed_Retrigger mode, i.e. every trigger event causes wave-
form output to restart.
BB:ARB:TRIG:EXEC
executes a trigger, waveform output is started.
BB:ARB:TRIG:ARM:EXEC
waveform output is stopped.
BB:ARB:TRIG:EXEC
executes a trigger, waveform output is started again.
Usage: Event
[:SOURce<hw>]:BB:ARB:TRIGger:EXECute
The command executes a trigger. The internal trigger source must be selected using the
command ARB:TRIGger:SOURce INTernal and a trigger mode other than AUTO must
be selected using the command :ARB:SEQuence.
Example:
BB:ARB:TRIG:SOUR INT
sets internal triggering.
BB:ARB:SEQ RETR
sets Retrigger mode, i.e. every trigger event causes signal output
to restart.
BB:ARB:TRIG:EXEC
executes a trigger.
Usage: Event
[:SOURce<hw>]:BB:ARB:TRIGger:EXTernal:SYNChronize:OUTPut <Output>
(enabled for "Trigger Source" External)
Enables/disables output of the signal synchronous to the external trigger event.
For or two or more R&S SMBVs configured to work in a master-slave mode for synchro-
nous signal generation:
●
disable this parameter in the slave instruments, in case the master instrument pro-
vides the slaves with its internal trigger signal and
●
use the default enabled state, if a common external trigger event is provided for the
master and the slave instruments.
SOURce Subsystem