Remote Control Commands
R&S
®
SMBV100A
512Operating Manual 1407.6062.32 ─ 08
Example:
BB:ARB:TRIG:SOUR EXT
sets external triggering via the TRIGGER 1 connector.
BB:ARB:TRIG:MODE ARET
selects the Armed_Retrigger mode.
BB:ARB:TRIG:RMOD?
queries the current status of waveform output.
Response: RUN
the waveform is output, an external trigger was executed.
Usage: Query only
[:
SOURce<hw>]:BB:ARB:TRIGger:SLENgth <SLength>
The command defines the length of the signal sequence to be output in the Single trigger
mode. The unit is defined with command SOUR:BB:ARB:TRIG:SLUNit. It is possible
to output deliberately just part of the waveform, an exact sequence of the waveform, or
a defined number of repetitions of the waveform.
Parameters:
<SLength> float
Range: 1 to 2^32-1
*RST: 1 Waveform length
Default unit: sample
Example:
BB:ARB:SEQ SING
sets trigger mode Single.
BB:ARB:TRIG:SLUN SAMP
sets unit Samples for the entry of sequence length.
BB:ARB:TRIG:SLEN 200
sets a sequence length of 200 samples. The first 200 samples of
the current waveform will be output after the next trigger event.
[:SOURce<hw>]:BB:ARB:TRIGger:SLUNit <SLunit>
The command defines the unit for the entry of the length of the signal sequence
(SOUR:BB:ARB:TRIG:SLEN) to be output in the "Single" trigger mode
(SOUR:BB:ARB:SEQ SING).
Parameters:
<SLunit> SEQuence | SAMPle
*RST: SEQuence
Example:
BB:ARB:SEQ SING
sets trigger mode Single.
BB:ARB:TRIG:SLUN SEQ
sets unit Sequence length for the entry of sequence length.
BB:ARB:TRIG:SLEN 2
sets a sequence length of 2 waveforms. The current waveforms
will be output twice after the next trigger event.