Remote Control Commands
R&S
®
SMBV100A
520Operating Manual 1407.6062.32 ─ 08
Parameters:
<Offset> float
Range: -1 to 1
Increment: 0.001
*RST: 0
Default unit: FS
Example:
BB:ARB:TSIG:RECT:OFFS 0.5
sets the DC component
[:SOURce<hw>]:BB:ARB:TSIGnal:RECTangle:SAMPles <Samples>
Sets the number of sample values required for the rectangular signal per period.
Parameters:
<Samples> integer
Range: 4 to 1000
Increment: 1
*RST: 100
Example:
BB:ARB:TSIG:RECT:SAMP 400
sets 400 samples per period
[:SOURce<hw>]:BB:ARB:TSIGnal:SINE:FREQuency <Frequency>
The command sets the frequency of the simple sinusoidal test signal. This signal is used
as output via the I channel. A sine wave of the same frequency but optionally phase-
shifted is generated on the Q path (ARB:TSIGnal:SINE:POFFset).
Parameters:
<Frequency> float
Range: 100 to Max
Increment: 0.01
*RST: 1000
Default unit: Hz
Example:
BB:ARB:TSIG:SINE:FREQ 100 kHz
sets a sine signal of 100 kHz.
[:SOURce<hw>]:BB:ARB:TSIGnal:SINE:PHASe <Phase>
The command sets the phase offset of the sine wave on the Q channel relative to the
sine wave on the I channel.
Parameters:
<Phase> float
Range: -180 to 180
Increment: 0.01
*RST: 90
Default unit: DEG
SOURce Subsystem