Remote Control Commands
R&S
®
SMBV100A
529Operating Manual 1407.6062.32 ─ 08
Parameters:
<Step> float
Range: -1 to 1
Increment: 1E-9
*RST: 0
Default unit: s
Example:
BB:ARB:MCAR:EDIT:CARR:DEL 5 us
sets a start delay of 5 us for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:DEL:STEP 1 us
the start delay is incremented by 1us for each carrier, i.e. the first
carrier has a start delay of 5us, the second a start delay of 6 us,
etc.
[:SOURce<hw>]:BB:ARB:MCARrier:EDIT:CARRier:DELay[:STARt] <Start>
The command sets the start delay for the individual carriers in the defined carrier range.
If the command :BB:ARB:MCAR:EDIT:CARR:DEL:STEP is used to define a step width,
the delay entered here applies only to the starting carrier. The delays of the remaining
carriers are stepped up or down by the delay value specified in
the :BB:ARB:MCAR:EDIT:CARR:DEL:STEP command.
Parameters:
<Start> float
Range: 0 to 1
Increment: 1E-9
*RST: 0
Default unit: s
Example:
BB:ARB:MCAR:EDIT:CARR:DEL 5us
sets a start delay of 5 us for the carriers in the carrier range.
[:SOURce<hw>]:BB:ARB:MCARrier:EDIT:CARRier:EXECute
The command adopts the settings for the carrier range which has been defined using
the :BB:ARB:MCAR:EDIT:CARR:... commands.
Example:
BB:ARB:MCAR:EDIT:CARR:STAR 4
the carrier range starts at carrier 2.
BB:ARB:MCAR:EDIT:CARR:STOP 20
the carrier range stops at carrier 20.
BB:ARB:MCAR:EDIT:CARR:STAT ON
sets all the carriers in the carrier range (2 to 20) to ON.
BB:ARB:MCAR:EDIT:CARR:EXEC
transfers the assistant settings for carrier 2 to 20 into the carrier
table.
Usage: Event
SOURce Subsystem