Remote Control Commands
R&S
®
SMBV100A
565Operating Manual 1407.6062.32 ─ 08
<RCosine> float
Range: 0.05 to 1.0
Increment: 0.01
*RST: 0.35
<SPhase> float
Range: 0.15 to 2.5
Increment: 0.01
*RST: 2
Example:
BB:DM:FILT:PAR:APCO25 0.2
sets the roll-off factor to 0.2 for filter type APCO25.
[:
SOURce<hw>]:BB:DM:FILTer:PARameter:COSine:BANDwidth <FiltParm>
Sets the bandwidth of the cosine filter.
Parameters:
<FiltParm> float
Range: 400 to 25E6
Increment: 1E-3
*RST: 270833.333
Example:
:BB:DM:FILT:TYPE COS
:BB:DM:FILT:PAR:COS:BAND 7200
[:SOURce<hw>]:BB:DM:FILTer:PARameter:APCO25Lsm:GAUSs <FiltParm>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:APCO25Lsm:LOWPass <FiltParm>
Sets the cut off frequency for the lowpass/ gauss filter.
Parameters:
<FiltParm> float
Range: 400 to 25E6
Increment: 1E-3
*RST: 270833.333
Example:
:BB:DM:FILT:TYPE APCO25Lsm
:BB:DM:FILT:PAR:APCO25L:LOWP 5166.667
:BB:DM:FILT:PAR:APCO25L:GAUS 1700
[:SOURce<hw>]:BB:DM:FILTer:TYPE <Type>
The command selects the filter type.
When a standard is selected (:BB:DM:STAN), the filter type and filter parameter are set
to the default value.