Remote Control Commands
R&S
®
SMBV100A
571Operating Manual 1407.6062.32 ─ 08
Parameters:
<Shape> LINear | COSine
*RST: COSine
Example:
BB:DM:PRAM:SHAP COS
selects a cosine-shaped rise and fall for the transmitted power
edge.
[:SOURce<hw>]:BB:DM:PRAMp:SOURce <Source>
The command sets the source for the power ramp control signals.
Parameters:
<Source> INTernal
INTernal
The internal control signals LEV_ATT and BURST are used. Con-
trol signals are stored in dedicated lists. When remote control is in
use, the list to be used is defined, selected and enabled with the
aid of SOURce:BB:DM:CLISt:... commands.
*RST: INTernal
Example:
BB:DM:PRAM:SOUR INT
enables the use of internal control signals for power ramp control.
[:SOURce<hw>]:BB:DM:PRAMp:TIME <Time>
The command sets the power ramping rise time and fall time for a burst.
Parameters:
<Time> float
Range: 0.25 to 16
Increment: 0.01
*RST: 1
Default unit: symbol
Example:
BB:DM:PRAM:TIME 2
sets a time of 2 symbols for the edges to rise and fall.
[:SOURce<hw>]:BB:DM:PRAMp[:STATe] <State>
The command enables or disables power ramping.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BB:DM:PRAM:STAT ON
switches power ramping on.
SOURce Subsystem