Remote Control Commands
R&S
®
SMBV100A
570Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:BB:DM:PRAMp:BBONly[:STATe] <State>
This command selects power ramping in the baseband only or mixed power ramping in
the baseband and the RF section.
The ON setting is mandatory if, with power ramping active, only the baseband signal is
output (I/Q outputs)
Only then can a signal with a defined, predictable level be output.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
BB:DM:PRAM:BBON ON
selects power ramping in the baseband only.
[:
SOURce<hw>]:BB:DM:PRAMp:FDELay <FDelay>
The command sets the delay in the falling edge. A positive value gives rise to a delay
and a negative value causes an advance. The setting is expressed in symbols.
Parameters:
<FDelay> float
Range: -4 to 4
Increment: 0.01
*RST: 0
Default unit: symbol
Example:
BB:DM:PRAM:FDEL 1
The falling edge starts 1 symbol later.
[:SOURce<hw>]:BB:DM:PRAMp:RDELay <RDelay>
The command sets the delay in the rising edge. A positive value gives rise to a delay and
a negative value causes an advance. The setting is expressed in symbols.
Parameters:
<RDelay> float
Range: -4 to 4
Increment: 0.01
*RST: 0
Default unit: symbol
Example:
BB:DM:PRAM:RDEL 1
The rising edge starts 1 symbol later.
[:SOURce<hw>]:BB:DM:PRAMp:SHAPe <Shape>
The command sets the edge shape of the ramp envelope.