Remote Control Commands
R&S
®
SMBV100A
607Operating Manual 1407.6062.32 ─ 08
Example:
BB:MCCW:TRIG:OUTP2:PATT #HE0F52,20
sets a bit pattern.
BB:MCCW:TRIG:OUTP2:MODE PATT
activates the marker signal according to a bit pattern on output
MARKER 2.
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:PULSe:DIVider <Divider>
The command sets the divider for Pulse marker mode
(SOUR:BB:MCCW:TRIG:OUTP:MODE PULSe). The resulting pulse frequency is derived
by dividing the symbol rate by the divider.
Parameters:
<Divider> integer
Range: 2 to 1024
Increment: 1
*RST: 2
Example:
BB:MCCW:TRIG:OUTP2:PULS:DIV 2
sets the divider for the marker signal on output MARKER 2 to the
value 2.
BB:MCCW:TRIG:OUTP2:FREQ?
queries the resulting pulse frequency of the marker signal
Response: 66 000
the resulting pulse frequency is 66 kHz.
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:PULSe:FREQuency?
The command queries the pulse frequency of the pulsed marker signal in the setting
SOUR:BB:MCCW:TRIG:OUTP:MODE PULS. The pulse frequency is derived by dividing
the symbol rate by the divider. The divider is defined with command
SOUR:BB:MCCW:TRIG:OUTP:PULS:DIV.
Return values:
<Frequency> float
Increment: 1E-3
Example:
BB:MCCW:TRIG:OUTP2:PULS:DIV 4
sets the divider for the marker signal on output MARKER 2 to the
value 4.
BB:MCCW:TRIG:OUTP2:MODE PULS
enables the pulsed marker signal
BB:MCCW:TRIG:OUTP2:PULS:FREQ?
queries the pulse frequency for the marker signal.
Response: "33 000"
the resulting pulse frequency is 33 kHz.
Usage: Query only
SOURce Subsystem