Clock Settings
This section lists the remote control commands, necessary to configure the clock.
The clock settings are available for R&S SMx and R&S AMU instruments only.
[:SOURce<hw>]:BB:MCCW:CLOCk:MODE......................................................................608
[:SOURce<hw>]:BB:MCCW:CLOCk:MULTiplier................................................................608
[:SOURce<hw>]:BB:MCCW:CLOCk:SOURce...................................................................609
[:SOURce<hw>]:BB:MCCW:CLOCk:SYNChronization:MODE............................................609
[:SOURce<hw>]:BB:MCCW:CLOCk:SYNChronization:EXECute.........................................610
[:SOURce<hw>]:BB:MCCW:CLOCk:MODE <Mode>
The command enters the type of externally supplied clock (:BB:MCCW:CLOCk:SOURce
EXTernal). When MSAMple is used, a multiple of the sample clock is supplied via the
CLOCK connector and the sample clock is derived internally from this. The multiplier is
entered with the command BB:MCCW:CLOCk:MULTiplier.
Parameters:
<Mode> SAMPle | MSAMple
*RST: SAMPle
Example:
BB:MCCW:CLOC:MODE SAMP
selects clock type "Sample", i.e. the supplied clock is a sample
clock.
[:SOURce<hw>]:BB:MCCW:CLOCk:MULTiplier <Multiplier>
The command specifies the multiplier for clock type "Multiple Samples"
(:BB:MCCW:CLOCk:MODE MSAM) in the case of an external clock source.
Parameters:
<Multiplier> integer
Range: 1 to 64
Increment: 1
*RST: 4
Example:
BB:MCCW:CLOC:SOUR EXT
selects the external clock source. The clock is supplied via the
CLOCK connector.
BB:MCCW:CLOC:MODE MSAM
selects clock type "Multiple Samples", i.e. the supplied clock has
a rate which is a multiple of the sample rate.
BB:MCCW:CLOC:MULT 12
the multiplier for the external clock rate is 12.