Remote Control Commands
R&S
®
SMBV100A
614Operating Manual 1407.6062.32 ─ 08
[:SOURce]:BBIN:POWer:RMS?
This command queries the rms level of the external digital baseband signal.
Return values:
<Rms> float
Range: -100 to 10
Increment: 0.01
*RST: 0
Example:
BBIN:POW:RMS
queries the estimated rms level.
Usage: Query only
[:
SOURce]:BBIN:SRATe:ACTual?
This command queries the sample rate of the external digital baseband signal.
Note: The maximum value is 90 MHz with BBinR Revision 02.xx and 100 MHz with BBinR
Revision 03.00 or higher.
Return values:
<Actual> float
Range: 400 Hz to 100 MHz
Example:
BBIN:SRAT
queries the sample rate value of the external digital baseband sig-
nal.
Usage: Query only
[:SOURce<hw>]:BBIN:SRATe:FIFO[:STATus]?
This command indicates the state of the signal transmission.
Return values:
<Status> OK | OFLow | URUN
OK
Resampling is working properly.
OFL = Overflow
The external sample rate frequency is too high which causes an
overflow in signal transmission.
URUN = Underrun
The external sample rate frequency is too low which causes an
underrun in signal transmission.
*RST: OK
Example:
BBIN:SRAT:FIFO:STAT
queries the FIFO Status of the sample rate.
Usage: Query only
SOURce Subsystem