Remote Control Commands
R&S
®
SMBV100A
613Operating Manual 1407.6062.32 ─ 08
[:SOURce]:BBIN:OLOad:HOLD:STATe?
This command queries an overload since the last reset for evaluating the measurement.
The Overload Hold state can be reset via the "Reset" command, or is reset automatically
by starting a new measurement ("Auto Level Set") or by setting new values (Crest Factor,
PEP, Level).
Return values:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:HOLD:STAT?
queries the overload state.
Response: 0
the AC/DC converter has not been overloaded.
Usage: Query only
[:SOURce]:BBIN:OLOad:STATe?
This command queries the current overload state of the A/D converter.
Return values:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:STAT?
queries the overload state.
Response: 0
The AC/DC converter is not overloaded.
Usage: Query only
[:SOURce]:BBIN:POWer:PEAK <Peak>
This command enters the peak level of the external baseband signal relative to full scale
of 0.5 V (in terms of dB full scale).
Parameters:
<Peak> float
Range: -60 to 3.02
Increment: 0.01
*RST: 0
Default unit: dBfs
Example:
BBIN:POW:PEAK -2
enters the peak level of -2 dBfs.
SOURce Subsystem