Remote Control Commands
R&S
®
SMBV100A
686Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:PULM:DOUBle:STATe <State>
The command enables/disables double pulse generation. The two pulses are generated
in one pulse period.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
PULM:DOUB:STAT ON
double-pulse mode is enabled.
Options: R&S SMBV-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:DOUBle:WIDTh <Width>
The command sets the width of the second pulse in case of double pulse generation.
Parameters:
<Width> float
Range: 10 ns to 100 s
Increment: 10 ns
*RST: 3 us
Default unit: ns
Example:
PULM:DOUB:WIDT 33 us
sets a width of 33 us for the second pulse.
Options: R&S SMBV-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:MODE <Mode>
Sets the mode of the pulse generator.
Parameters:
<Mode> SINGle | DOUBle
SINGle
Enables single pulse generation.
DOUBle
Enables double pulse generation. The two pulses are generated
in one pulse period.
*RST: SINGle
Example:
PULM:MODE DOUB
enables double pulse generation.
Options: R&S SMBV-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:OUTPut:SYNC[:STATe] <Sync>
Configures the signal at the SIGNAL VALID connector.