Remote Control Commands
R&S
®
SMBV100A
688Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:PULM:SOURce <Source>
The command selects the source for pulse modulation.
Parameters:
<Source> INTernal | EXTernal
INTernal
The internally generated rectangular signal is used for the pulse
modulation. The frequency of the internal signal can be set in the
SOURce:LFOutput subsystem.
EXTernal
The signal applied externally via the EXT MOD connector is used
for the pulse modulation.
*RST: INTernal
Example:
PULM:SOUR INT
selects the internal modulation source.
Example:
PULM:STAT ON
activates the pulse modulation.
Usage: SCPI confirmed
Options: R&S SMBV-K22 (Pulse Modulator)
[:SOURce<hw>]:PULM:STATe <State>
The command activates/deactivates the pulse modulation.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
PULM:STAT ON
activates pulse modulation with the modulation source selected
under SOURce:PULM:SOURce.
Usage: SCPI confirmed
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity <Polarity>
Selects the polarity of the Gate signal.
The signal is supplied via the PULSE EXT connector.
Parameters:
<Polarity> NORMal | INVerted
*RST: NORMal
Example:
PULM:TRIG:EXT:GATE:POL NORM
The pulse signal is generated while the gate signal is high.
Options: R&S SMBV-K23 (Pulse Generator)
SOURce Subsystem