Remote Control Commands
R&S
®
SMBV100A
695Operating Manual 1407.6062.32 ─ 08
Parameters:
<Mode> AUTO | MANual | STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. Each frequency step of the sweep
is triggered individually, either by varying the "Current Fre-
quency" value using the rotary knob under manual control or by
means of a FREQ:MAN command under remote control. With man-
ual control, the frequency increases or decreases (depending on
the direction of the rotary encoder) by the value specified under
FREQ:STEP:INCRement. With remote control, the frequency is
set directly with the command :FREQ:MAN.
STEP
Each trigger triggers one sweep step only (Mode Single Step). The
frequency increases by the value entered under
SOUR:SWE:FREQ:STEP:LIN (linear spacing)
or ...:STEP:LOG (logarithmic spacing).
*RST:
AUTO
Example:
SWE:MODE AUTO
selects Mode Auto for a frequency sweep at the RF output.
[:SOURce<hw>]:SWEep[:FREQuency]:POINts <Points>
The command sets the number of steps in an RF sweep.
The command is linked to the command :SWEep[:FREQuency]:STEP as follows:
for linear sweeps
POINts = (SPAN / STEP:LIN) + 1
logarithmic sweeps and STARt < STOP:
POINts = ((log STOP - log STARt) / log STEP:LOG) + 1
If POINTs changes, the value of STEP is adjusted. The STARt and STOP value is retained.
Two separate POINts values are used for linear or logarithmic sweep spacing
(:SWEep[:FREQuency]:SPACing LIN | LOG). The command is always effective for
the currently set sweep spacing.
Parameters:
<Points> float
SOURce Subsystem