Sirona Dental Systems GmbH 4Operation
Operating Instructions INTEGO 4.5Dentist element
båÖäáëÜ
4.5.10.3
Beeps
In addition to the graphical distance display on the touchscreen, the
position of the file in the root canal can also be indicated as an acoustic
signal.
Apex acoustic signals
The ApexLocator always plays the following acoustic signals:
One beep sounds when the physiological apex is reached and the
motor stops automatically once it has reached the preset motor stop
position.
Three beeps sound if the motor is switched to counterclockwise
rotation when the auto-reverse function is activated and the foot
pedal is pressed again.
For manual measurements using the file clamp in the
Start
sub-screen,
Apex signal tones are not output.
Apex distance acoustic signals
If the
Apex distance signal tones
key is highlighted in orange, the
following signal tones are issued in addition to the Apex signal tones:
No acoustic signal is played if the file is at least five display levels (on
the distance display) away from the apex.
Beeps with long intervals are played if the file is three or four display
levels away from the apex.
Beeps with short intervals are played if the file is one or two display
levels away from the apex.
A constant signal is played when the file has reached or passed the
apex.
If both types of acoustic signal are switched on simultaneously, the
continuous signal sounds when the motor automatically stops after it
reaches the apex and the motor stop function is switched on. The three
beeps continue to sound if auto-reverse is switched on.