5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.4Vacuum system Operating Instructions INTEGO
Rinsing the cleaning agent
1. Following the cleaning process, pour at least 1 liter of water into the
cuspidor.
The water is drawn off, thus preventing any cleaning agent
residues from remaining in the suction hoses.
2. When the aspiration process has been completed, detach the hoses.
Place the suction hoses back in their holders.
3. Remove the cleaning adapter B and insert the gold trap A.
If the treatment center is equipped with a third suction hose, repeat the
above procedure.
Clean the filter insert of the standard wet suction and the receiving tank of the air jet pump
If the water unit is equipped with a standard wet suction, the filter insert
of the standard wet suction must also be cleaned once a month after
cleaning the suction system; see "Cleaning the filter insert of the standard
wet suction" [ 263].
If the Compact water unit is equipped with an air jet pump (Air Venturi),
the collector should be emptied weekly after cleaning the suction system,
see "Emptying the receiving tank of the air jet pump" [ 266].
5.4.3.2
Cleaning the suction system via external container
If the treatment center is not equipped with a cuspidor, the suction system
must be cleaned via external container.
1. Prepare 1 liter of cleaning solution in a separate container according
to the manufacturer's instructions and mix it thoroughly.
2. Fill the cleaning solution into a suitable container.
3. If the container is equipped with suitable adapters for the suction
hoses, take the suction cannulae from the suction hoses. Otherwise,
the cleaning solution is to be aspirated with plugged on suction
cannulas.
4. Set maximum suction flow on the suction handpieces.
5. Remove the suction hoses from their holders and aspirate the
cleaning solution simultaneously with all the suction hoses from the
container.
6. Allow the cleaning solution to act. Observe the reaction time specified
for the cleaning solution by the manufacturer.
7. Following the cleaning process, pour at least 1 liter of water into the
container. Aspirate the water in the same way to ensure that no
cleaning agent residues remain in the suction hoses.
8. After finishing, place the suction hoses back in their holders.