64 57 498 D3543
D3543.201.01.05.02 08.2016
45
Sirona Dental Systems GmbH 3Unit description
Operating Instructions INTEGO 3.9Dentist element
båÖäáëÜ
3.9.5.2
Fixed keys of the EasyTouch user interface
Clean key
Pressing this key deactivates the complete user interface of the dentist
element. Pressing it again > 3 s reactivates the user interface.
Deactivates the complete user interface of the dentist element. The
instruments of the dentist and assistant elements can no longer be
activated. The SIVISION monitor and the camera system are switched
off. Pressing it again (>3 s) reactivates the user interface. This is used to
clean the surface and protect against interference from an external HF
surgical unit so that no unwanted functions can be accidentally triggered,
see “Disinfecting user interfaces” [ → 209]. Electromagnetic disturbances
can often be reduced, if the external HF surgical unit is operated with a
neutral electrode. As long as the suction removal system is required with
the external HF surgical unit during the treatment, the suction handpiece
must be removed from the holder before blocking the treatment center.
The suction unit remains switched on until the block is removed and the
suction handpiece is put down again.
Tumbler filling
Starts or stops the tumbler filling function.
Pressing the
tumbler filling key
(> 2 s) displays the setting text for
coupling the tumbler filling to mouth rinsing position S and filling time, see
"Tumbler filling on the EasyTouch" [ → 149].
Flushing
Starts or stops cuspidor flushing.
Pressing the
flushing
key (> 2 s) displays the setting text for coupling
flushing to mouth rinsing position S and flushing time, see "Flushing of the
cuspidor on the EasyTouch" [ → 152].
Operating light
Switches the operating light on, to the composite function, or off.
The composite function is used to prevent premature curing of composite
materials.
Operating light setting program
With the LEDlight, light intensity is set using the contactless sensor. With
the LEDview and LEDview
PLUS
, when the
Operating light
key is pressed
(> 2 s), the operating context for the light appears. The temperature of the
light can be set with LEDview
PLUS
.
For details, see section "Operating light" [ → 169].
Sub-screen
Some programs are divided into a main program and sub-screens.
Only function keys for the basic functions are displayed in the main
dialogs. The
Sub-dialog
key (two rectangles) leads to further setting
possibilities.