64 57 498 D3543
D3543.201.01.05.02 08.2016
127
Sirona Dental Systems GmbH 4Operation
Operating Instructions INTEGO 4.5Dentist element
båÖäáëÜ
4.5.10.5
Endodontic treatment with ApexLocator and torque limitation
handpiece
If your treatment center is equipped with the ApexLocator option but does
not have integrated torque control, the Apex function can be used in the
Motor dialog for endodontic treatments.
The ApexLocator is available only with the EasyTouch Comfort user
interface.
ApexLocator, SIRONiTi APEX
The treatment center is prepared for apex measurements using the
instrument, see section "Preparing to use the ApexLocator".
The
Motor dialog
is displayed on the touchscreen.
1. Set the appropriate motor speed according to the handpiece and the
root canal file used, see "Setting the speed on the EasyTouch"
[ 102].
2. You can set the motor so that it stops automatically when the
physiological apex is reached. If you want to use the automatic motor
stop, this can be preset in the
Motor
sub-dialog. To do this touch the
Sub-dialog
key.
The
Motor
sub-dialog is displayed.
3. Touch the
Apex stop
key.
If the key is highlighted orange, the motor stops automatically
when the physiological apex is reached. The – and + keys and
the
AutoReverse
key are displayed.
4. Use the – and + keys to set the required apex distance from 1.5 to 0.
If the distance is 0, the motor does not stop until it reaches the
physiological apex. Please note that the distance values are not a
metric length measurement!
IMPORTANT
SIRONiTi APEX
Only Sirona SIRONiTi APEX must be used to perform apex
measurement using a torque-limiting handpiece! The electrical
conductivity can only be ensured using this handpiece.
The SIRONiTi handpiece can also be used to perform endodontic
treatments without the ApexLocator.
Auto
Rev
Apex
Stop
Apex
Stop
Auto
Rev
Apex
Stop
0.5
Apex