4Operation Sirona Dental Systems GmbH
4.6Assistant element Operating Instructions INTEGO
Quick curing mode
In the quick curing mode, the Mini L.E.D. operates at full power for 10
seconds.
In this mode, the light intensity output is as follows:
1250 mW/cm
²
(± 10%) with the standard light guide, dia. 7.5 mm
2000 mW/cm
²
(± 10%) with the booster light guide, dia. 5.5 mm
Pulsed curing mode
In the pulsed curing mode, the Mini L.E.D. radiates in 10 consecutive light
pulses of 1 s each. There is a 250 ms break between the individual
pulses.
"Soft start" mode
The “soft start” mode features:
A 10-second “soft start” from 0 to 1250 mW/cm
²
, or from 0 to 2000
mW/cm
²
with the "booster light guide," dia. 5.5 mm.
Full power for 10 seconds.
4.6.6.5
Operating the Mini L.E.D.
Curing light, contact with the material to be cured
When the Mini L.E.D. is removed from its holder, the operating mode last
used before it was deposited is selected.
1. Use the
Mode
key H to select the quick curing, pulse curing or "soft
start" mode.
The corresponding lamp shows which mode is selected. The Mini
L.E.D. is ready for operation.
2. Hold the light guide as close as possible to the composite material
surface to be photopolymerized.
3. Start the curing cycle. Press the on/off button C briefly.
An acoustic signal sounds. The curing cycle is started.
The acoustic signal sounds every 5 seconds.
The end of the curing cycle is also indicated by an acoustic signal.
You can interrupt the curing cycle immediately by pressing the on/off
button C.
For care and cleaning, see "Disinfecting and sterilizing the Mini L.E.D.
curing light" [ 232].
0510
Sec.
0
12.5
Sec.
051015 20
Sec.
IMPORTANT
Contact with the material to be cured
Make sure that the light guide never touches the material to be cured,
as this may cause damage to the light guide and reduce its
effectiveness.