Sirona Intego - Switching the Treatment Center On;Off

Sirona Intego
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 57 498 D3543
60 D3543.201.01.05.02 08.2016
4Operation Sirona Dental Systems GmbH
4.1Starting up the treatment center Operating Instructions INTEGO
4.1.2 Switching the treatment center on/off
The treatment center has a main switch with integrated fuse on the base
of the chair.
At the end of the working day, the treatment center should be switched off
to save energy and for safety reasons. This switches off the air and water
supply as well as all electronic components. The treatment center then
consumes no power.
Following switch-on, the operating system is booted and an automatic
self-test is performed.
Switching the treatment center on
The treatment center is installed by authorized technical personnel
according to the "Installation Instructions".
Turn on mains switch A.
The treatment center is connected to the power grid and air and water
supply.
The treatment center powers up and establishes operational
readiness.
Self-test for chair drive
After switching on the treatment center, an automatic self-test for the
chair drive is performed. The patient chair moves up and down briefly.
The test is only performed at the start of the day or if the chair has not
been moved for longer than 6 hours.
Patient chair, self-test for chair drive
Next maintenance call
If the next maintenance call is due in less than 42 days or the
maintenance deadline has already been exceeded, a message appears
on the user interface. For more information, please refer to "Inspection
and maintenance" [ 292].
Switching the treatment center off
Turn off mains switch A.
The treatment center is disconnected from the power supply. The air
and water supply is blocked.
A
CAUTION
During this self-test, there must not be a patient on the chair.
If a safety switch has been triggered, the self-test will not be performed.
The test, however, starts immediately after the cause has been
remedied, e.g., after swiveling the cuspidor outwards. There is a risk of
collision for the patient.
Do not allow anyone to sit on the chair until the treatment center has
been switched on and is ready for use.