Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.4Vacuum system
båÖäáëÜ
5.4.2.2
Suction hose cleaning on the Comfort water unit
The Comfort water unit is always equipped with an adapter for the suction
hose cleaning function. This can be used to purge the suction hoses
consecutively. A cleaning agent is automatically added to the water if the
dental treatment center is equipped with the chemical suction hose
cleaning option.
Cleaning suction hoses and suction system
1. Remove the suction tips from the suction hoses.
2. Set maximum suction flow on the suction handpiece to be cleaned.
All the other suction handpieces must be completely closed.
3. Swivel the cover cap on the assistant's side of the water unit upward
for fitting the suction hoses.
4. Attach the open suction handpiece to the receptacle.
5. Press the button.
The mixture of water and cleaning agent (optional) is pumped
into the suction hose cleaning container and drawn off by the
suction hoses. An acoustic signal will sound after completing the
suction hose cleaning.
6. Then place the suction hoses back into their holders on the assistant
element.
7. Repeat this procedure for the other suction handpieces.
Vacuum system, hose fittings
CAUTION
Suction hose cleaning must not be interrupted to ensure that no residual
water remains in the container.
CAUTION
Disinfect the hose fittings on the water unit after each patient.