5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.3Instruments and instrument hoses Operating Instructions INTEGO
5.3.5 Lubricate, disinfect/sterilize the treatment instruments
5.3.5.1
Treatment instruments
The procedures required for the following treatment instruments are
described in the following separate operating instructions:
List of treatment instruments, E motor, without implant and HF surgery
● BL E motor
● BL ISO E motor
● Straight and contra-angle handpieces in various versions
● Highspeed handpieces
● 3-way syringeSPRAYVIT E
● Standard3-way syringe, see "Cleaning and sterilizing the standard 3-
way syringe" [ → 230]
● SIROSONIC L Scaler
5.3.5.2
Cleaning and sterilizing the standard 3-way syringe
The Standard 3-way syringe can be cleaned under running water. Spray
and wipe disinfection can be used. The jet can be removed from the tip
for sterilization.
➢ Press on the ring-shaped unlock button.
The jet adapter is released. The jet will fall out.
To attach the jet, press it in until it clicks audibly into place.
Multi-purpose syringe, nozzle falling out
WARNING
The nozzle of the Standard 3-way syringe must lock into the holder.
Otherwise, the jet could fall out during treatment.
➢ Prior to treatment, ensure that the jet is securely in place.