Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.3Instruments and instrument hoses
båÖäáëÜ
5. Activate the spray.
Toggle the switch on the pneumatic foot switch to the right.
Activate the left key on the C
+
electronic foot control.
6. Set a low speed and intensity for the instrument. Press favorite key 1.
7. Press and hold the foot pedal for at least 60 seconds.
8. Afterward, return the instrument to the instrument holder.
9. Repeat the procedure for all other instruments.
The manual purging process is finished. The treatment center is
again ready for operation.