64 57 498 D3543
242 D3543.201.01.05.02 08.2016
5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.4Vacuum system Operating Instructions INTEGO
5.4.3 Cleaning vacuum system using cleaning adapter in the
cuspidor or via external container
Compact and Comfort water unit
If the Compact water unit is not equipped with the suction hose cleaning
option, the suction system must be cleaned daily using the cleaning
adapter in the cuspidor or via external container.
If the Comfort water unit is not equipped with the chemical suction hose
cleaning option, only water is used for cleaning the suction hoses. The
suction system therefore must be cleaned daily using the cleaning
adapter in the cuspidor or via external container.
The suction system is exposed to secretions, saliva, and blood that
contain bacteria. Cleaning at regular intervals is therefore absolutely
mandatory for hygienic reasons.
Care, cleaning, and disinfecting agents
Vacuum system, foaming cleaning agents
5.4.3.1
Cleaning vacuum system using cleaning adapter in the cuspidor
Preparation for cleaning
1. Prepare 1 liter of cleaning solution in a separate container according
to the manufacturer's instructions and mix it thoroughly.
2. Remove the gold trap A.
3. Clean the cuspidor.
4. Insert the cleaning adapter B as far as it will go.
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ 205].
NOTICE
Domestic cleaning agents foam up.
Foaming cleaning agents often cause foam and water to be sucked into
the dry suction system. This can cause damage to the suction machine.
Use only care, cleaning and disinfecting agents approved by Sirona,
see"Care, cleaning, and disinfecting agents" [ 205].
B
A
B
A