Sirona Dental Systems GmbH 4Operation
Operating Instructions INTEGO 4.4Patient chair
båÖäáëÜ
4.4.10 Programming chair programs
The four chair programs preset in the factory:
● Mouth rinsing position S
● Entry/exit position 0
● 1 and 2
can be individually programmed for both user profiles (A and B) on the
EasyPad and each of the four user profiles (A to D) on the EasyTouch.
✔ On the EasyTouch, the
Start
or
Instrument dialog
is displayed on the
touchscreen.
1. Move the patient chair to the required treatment position; see "Moving
the chair manually" [ → 77].
2. Switch the operating light on or off (this will be included in the
program), see “Operating light” [ → 169].
3. Press and hold the desired program key (S, 0, 1, 2) (> 2 s).
An acoustic signal sounds. Your settings are now stored under
the desired program key.
Note: Chair programs S and 0 can also be programmed on the assistant
element side.