EasyManua.ls Logo

Sirona Intego - Disinfect the Tray

Sirona Intego
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 57 498 D3543
D3543.201.01.05.02 08.2016
211
Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.2Surfaces
båÖäáëÜ
Assistant element
The silicone mat of the Comfort assistant element is identical with that of
the CS dentist element.
1. Switch the treatment center off and remove the suction hoses and
instruments from the holders of the assistant element.
2. Clean and disinfect the assistant element, see "Cleaning and
disinfecting surfaces" [ 208].
3. For the Comfort assistant element, place the sterilized silicone mat on
the assistant element.
4. Place the suction hoses and instruments back in their holders.
5.2.4 Disinfect the tray
The tray can be removed to facilitate cleaning or thermal disinfection.
Care, cleaning, and disinfecting agents
1. Hold the tray tightly.
2. Open the lock by swiveling the lever downwards.
3. Remove the tray.
4. Let the lever fall back into its original position.
5. Thermally disinfect the tray.
To insert the tray, simply guide it into the mount. The mechanism locks
automatically.
Tray, Tra y Holder
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ 205].
93°C
200°F
93°C
200°F
CAUTION
If the tray is not locked in place, it can disengage from the tray holder.
After installing the tray, make sure it is securely attached to the tray
holder.

Table of Contents

Other manuals for Sirona Intego

Related product manuals