Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.6Sanitizing
båÖäáëÜ
5.6.2 Sanitizing with disinfection system
Sanitizing must be performed:
Regularly every 4 weeks when operating with the public drinking
water supply or if stand-alone water supply is used more than 28 days
in exceptional cases
If the treatment center features a disinfection system, the display
Days until next sanitation
appears on the EasyPad or touchscreen. It
first appears three days before the date of sanitation.
After longer periods of disuse (> one week)
If the germination index exceeds 100 germs per milliliter, see
"Microbiological water test" [ 205].
After changing from stand-alone water supply to public drinking water
supply and DENTOSEPT P disinfectant
The sanitizing process takes at least 24 hours.
Additional devices, sanitation with disinfectant
Preparation
The following preparations should be made prior to beginning sanitation.
1. If your treatment center is equipped with a cuspidor, activate the
cuspidor flushing for at least one minute. This flushes the water lines.
2. Set all burr drives and the SIROSONIC L scaler to the maximum
water flow rate. The water regulators under the instrument holder
must be turned counterclockwise.
3. Put all instruments and suction hoses in place.
4. Do not remove the tumbler holder from the cuspidor. Put an empty
glass with a volume 200 ml under the tumbler outlet to prevent
discoloration from the DENTOSEPT P disinfectant.
41d
B
10:45:03
San
3d
NOTICE
Additional devices connected to the external device connection must
not be sanitized with the treatment center.
Additional devices can be damaged. Residues of DENTOSEPT P
disinfectant can remain in the additional devices.
Disconnect any additional devices from the treatment center prior to
sanitation.
max.
IMPORTANT
The water flow rate to the instruments is not checked by the treatment
center.
Make sure that the maximum water flow rate is set for all
instruments to be purged.