Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.5Components of the water unit
båÖäáëÜ
5.5.9 Cleaning the filter insert of the wet suction device with
cuspidor valve
The automatic separator and amalgam separator / sediment container or
air jet pump are not installed in the water unit for standard wet suction.
The separation of air and water and amalgam separation are performed
centrally in this case.
In order to ensure that larger solid particles nevertheless cannot enter the
separating unit, the suction line in the water unit is equipped with a filter.
The filter insert must be cleaned when the suction power decreases.
Wear gloves, risk of contamination
Cleaning the suction system
Before the filter insert of the standard wet suction is cleaned, the suction
system should be cleaned
If the treatment center is equipped with the chemical suction hose
cleaning option, see "Cleaning the suction hoses" [ → 235]. If this option
is not available, see "Cleaning vacuum system using cleaning adapter in
the cuspidor or via external container" [ → 242].
Suctioning off remaining water completely, removing and cleaning the
filter insert
As long as the treatment center is switched on, water remains in the
suction line for technical reasons. Therefore, in order to clean the filter
insert, this water must be completely extracted beforehand. Otherwise
the remaining water will flow out when the filter housing is opened.
Wet suction
WARNING
Wear gloves when performing the following work.
NOTICE
The flushing and tumbler filling functions may not be activated when the
filter housing is open.
Otherwise, water could escape from the open filter housing.
➢ Do not switch the flushing and tumbler filling functions on whenever
the filter housing is open.