Sirona Dental Systems GmbH 4Operation
Operating Instructions INTEGO 4.4Patient chair
båÖäáëÜ
4.4
Patient chair
4.4.1 Safety instructions
Patient chair, chair movements
Maximum load capacity of patient chair
Patient chair, movement range
Patient chair, interlocked when instrument is active
If chair movement is permanently blocked, please contact your service
technician.
ü û
CAUTION
The free space under the patient couch and up to the water unit can be
decreased due to chair movements.
Parts of the patient's or user's body may be pinched or crushed.
Do not allow any limbs to stick out in the space between the chair
upholstery, armrests and chair base. Please make sure that the
patient’s arms and legs rest on the upholstery of the chair.
Do not place any objects on the base of the chair.
max. load of chair
185 kg
Intermittent operation
of chair motors:
2 min ON/18 min OFF
max. load of chair
140 kg
Intermittent operation
of chair motors:
2 min ON/18 min OFF
CAUTION
The maximum load capacity of the patient chair is 140 kg (308.6 lbs) or
185 kg (407.9 lbs), depending on the model (tested with multiple safety
acc. to IEC 60601-1).
If the maximum load capacity is exceeded, there is a risk of damage to
the treatment chair and injury of the patient.
Never allow any persons who weigh more than 135 kg (297.6 lbs)
or 180 kg (396.8 lbs) to sit on the patient chair. The permissible
maximum load capacity is indicated on a plate next to the rating
plate of the treatment center.
The maximum additional weight of accessories mounted on the
patient chair is 5 kg (11 lbs).
CAUTION
Objects may protrude into the movement range of the chair.
There is a risk of crushing the patient and damaging the objects.
Make sure that no objects such as e.g. windows, drawers or other
devices protrude into the movement range of the treatment chair.
IMPORTANT
Chair interlock
As long as a treatment instrument is activated, all functions for moving
the patient chair are disabled for safety reasons.