64 57 498 D3543
206 D3543.201.01.05.02 08.2016
5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.1Basics Operating Instructions INTEGO
5. Immerse the tester in the filled container for 30 seconds.
The cardboard disk with the culture medium will now absorb 1 ml
of the water sample.
6. Remove the tester from the container. Shake out any excess water.
Empty the tank.
7. Place the tester in the container for incubation either for two days at
a room temperature of 20°C/68°F or for 24 hours at a temperature of
35°C/95°F.
8. Count all colonies found on the surface of the tester.
If the number of germs is well over 100, the water must be treated,
see"Sanitizing" [ → 268] and "Biofilm removal by the service technician"
[ → 290].
5.1.4 General handling instructions
The general handling instructions mainly apply to the treatment center
provided there are no other product-specific handling instructions in this
operating manual. The manufacturer's instructions on the disinfecting
agents should be followed (temperature, concentration, application times,
etc.).
Manual cleaning
Manual cleaning can be done with a cloth or soft brush. Unless specified
otherwise, use lukewarm drinking water for cleaning surface dirt.
Machine cleaning
Thermal disinfection of up to 93 °C according to ISO 15883-1 is possible
for components marked accordingly. Use an alkaline cleaning agent for
this purpose. A laboratory washing machine, such as the Miele G7781,
may be used as a cleaning device. The setting of the device as well as
the type and amount of cleaning agent can be found in the operating
manual of the respective machine.
Manual disinfection
The treatment center can be disinfected by wiping. Use a soft, colorless
cloth for this as well as an approved disinfectant. Unless specified
otherwise in the operating manual, other disinfection methods (spray
disinfection, immersion, etc.) may not be used. Please observe the
information regarding the use of disinfectants in the treatment center
operating manual (temperature, concentration, application times, etc.).
48 h – 20 °C / 68 °F
24 h – 35 °C / 95 °F