Sirona Dental Systems GmbH 4Operation
Operating Instructions INTEGO 4.1Starting up the treatment center
båÖäáëÜ
4
Operation
4.1
Starting up the treatment center
4.1.1 Initial Operation
Sanitizing must be performed prior to initial startup of your treatment
center if the treatment center is equipped with this option.
In the INTEGO, the purging and sanitizing functions and the fresh water
bottle are available as options. Both options must be present to sanitize
the water lines. If it is not possible to sanitize the treatment center, the
water lines of the instrument and tumbler filling must be flushed manually
before the initial start-up, see "Switch spray on/off" [ → 87] and "Purge
water lines" [ → 214].
In the INTEGO pro, the sanitize function is present provided that the
water unit is equipped with a disinfection system.
For sanitizing, all water-bearing lines are filled with DENTOSEPT P
concentrated disinfectant to reduce the bacteria count of the water.
If the service engineer skipped the sanitizing procedure after installing
your treatment center based on an agreement with you or sanitizing has
not been performed for more than one week, please carry out sanitizing
yourself, see "Sanitizing" [ → 268].
Sanitizing takes 24 hours.