EasyManua.ls Logo

Sirona Intego - Components of the water unit

Sirona Intego
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 57 498 D3543
248 D3543.201.01.05.02 08.2016
5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.5Components of the water unit Operating Instructions INTEGO
5.5
Components of the water unit
5.5.1 Clean the gold trap
The gold trap retains larger solid particles to prevent them from being
washed down the drain in the cuspidor bowl. This means that the
amalgam rotor must be replaced less often and the sediment container or
collector of the air jet pump must be emptied less often.
Amalgam residues, amalgam rotor, sediment container, or collec tor of the air jet pump
1. Remove the gold trap from the drain of the cuspidor bowl.
2. Remove the amalgam residues from the gold trap. Amalgam residues
must be disposed of separately.
3. Clean the gold trap.
4. Reinsert the gold trap.
CAUTION
Amalgam residues must not enter the public sewage system.
Amalgam is a mercury compound that is hazardous to water.
Do not dispose of amalgam residues in a sink.
Collect amalgam residues in a closed container with water. Dispose
of amalgam residues, e.g. when replacing the amalgam rotor, by
filling the amalgam residues into the amalgam rotor or when
emptying the sediment container or collector of the air jet pump.

Table of Contents

Other manuals for Sirona Intego

Related product manuals