Sirona Dental Systems GmbH 4Operation
Operating Instructions INTEGO 4.7Water unit
båÖäáëÜ
4.7.5 Converting the Ambidextrous water unit for right and left-
handed people
Preparing the treatment center
1. If an electric foot switch is available; bring the electric foot switch out
of the swiveling area of the water unit.
CAUTION
There must be no patient in the chair during conversion.
The patient would otherwise collide with treatment center components.
Move the patient chair to the entry/exit position 0 and guide the
patient out of the movement range.
CAUTION
To prevent injuries caused by falling objects, never place or hang
anything on the dentist or assistant element, the tray or its support arms
during conversion.
NOTICE
During conversion the water unit with the dentist element, assistant
element and operating light are moved on the opposite side of the
treatment center.
Hereby components of the water unit could collide with obstacles such
as walls, furniture and windows.
Observe the following instructions to convert the treatment center
with the smallest possible space requirements.
Always move the components with extreme care.