4Operation Sirona Dental Systems GmbH
4.7Water unit Operating Instructions INTEGO
In the
Start
sub-screen, the
Aqua
key is highlighted in orange.
The disinfection system is now changed to stand-alone water
supply.
Resetting disinfection system for operation with public drinking water
and disinfectants
Resetting the disinfection system to public drinking water is similar to
changing to stand-alone water supply. Instead of the
Refilling water
display, the
Refilling disinfectant
display appears on the touchscreen.
1. Fill approx. 0.5 liters of DENTOSEPT P into the disinfectant tank.
2. Sanitize the water lines after switching to the public drinking water
supply, see “Sanitizing with disinfection system” [ → 277].
Level control
If the
Refill water
or
Refill disinfectant
display appears during the
treatment, the reservoir is almost empty (< 300 ml). By pressing the
Return
key, the display can be hidden and the treatment can be
continued. Refill the tank as soon as possible.
Water unit, disinfectant tank < 300 ml
Setup
San
Aqua
Apex Dist
Auto
Setup
San
Aqua
Auto
Apex Dist
Stop
Desinf
IMPORTANT
300 ml are approximately sufficient for:
➢ 5 min. spray operation (motor, turbine, SIROSONIC L) or
➢ 2 min. SPRAYVIT operation or
➢ 4 x tumbler fillings
The actual values depend on the instrument types and settings used.