64 57 498 D3543
D3543.201.01.05.02 08.2016
25
Sirona Dental Systems GmbH 3Unit description
Operating Instructions INTEGO 3.2Technical data
båÖäáëÜ
3.2
Technical data
Model designation, power ratings
Unit classifications, without HF surgery
Mode of operation
Model designation: INTEGO / INTEGO pro
Power connection: 100 – 240 V AC ± 10%
50/60 Hz
Rated current: 3.3 A – 1.5 A at 100 – 240 V
also max. 6 A for external devices
Type of ground connection: TN-C-S system or TN-S system
(acc. to IEC 60364-1)
Overvoltage category: 2 acc. to IEC 60664-1
Average power consumption
(for dimensioning an air
conditioning system):
100 W
Power consumption when
switched off:
0 W (power switch present)
Main building fuse: Type B automatic circuit breaker
100 - 115 VAC: 20 A medium-blow
220 – 240 V AC: 16 A medium-blow
Protection class: Class I device
Device class in accordance
with Directive 93/42/EEC:
Class IIa equipment
Degree of protection against
electrical shock:
Type B applied parts
External intraoral camera SiroCam F /
AF / AF
+
. These are:
Applied part type BF
Degree of protection against
ingress of water:
Ordinary equipment (without protection
against ingress of water)
The foot switch has an IPX1 degree of
protection against liquids (drip-proof).
Mode of operation Continuous operation with intermittent
loading corresponding to the dental
mode of working. [ → 18]
Drive motors for chair operation:
intermittent use, max. 2 minutes on and
18 minutes off
Permanently connected unit. Operation
is not permitted in mobile vehicles.