64 57 498 D3543
D3543.201.01.05.02 08.2016
25
Sirona Dental Systems GmbH 3Unit description
Operating Instructions INTEGO 3.2Technical data
båÖäáëÜ
3.2
Technical data
Model designation, power ratings
Unit classifications, without HF surgery
Mode of operation
Model designation: INTEGO / INTEGO pro
Power connection: 100 240 V AC ± 10%
50/60 Hz
Rated current: 3.3 A – 1.5 A at 100 – 240 V
also max. 6 A for external devices
Type of ground connection: TN-C-S system or TN-S system
(acc. to IEC 60364-1)
Overvoltage category: 2 acc. to IEC 60664-1
Average power consumption
(for dimensioning an air
conditioning system):
100 W
Power consumption when
switched off:
0 W (power switch present)
Main building fuse: Type B automatic circuit breaker
100 - 115 VAC: 20 A medium-blow
220 240 V AC: 16 A medium-blow
Protection class: Class I device
Device class in accordance
with Directive 93/42/EEC:
Class IIa equipment
Degree of protection against
electrical shock:
Type B applied parts
External intraoral camera SiroCam F /
AF / AF
+
. These are:
Applied part type BF
Degree of protection against
ingress of water:
Ordinary equipment (without protection
against ingress of water)
The foot switch has an IPX1 degree of
protection against liquids (drip-proof).
Mode of operation Continuous operation with intermittent
loading corresponding to the dental
mode of working. [ 18]
Drive motors for chair operation:
intermittent use, max. 2 minutes on and
18 minutes off
Permanently connected unit. Operation
is not permitted in mobile vehicles.