3.2Technical data Operating Instructions INTEGO
Operating and transport conditions
Load capacity of the patient chair
Approval and year of manufacture
Minimum PC requirements
Transport and storage
conditions:
Temperature: -40 °C – +70 °C
(-40 °F – 158 °F)
Relative humidity: 10% – 95%
Air pressure: 500 hPa – 1,060 hPa
Operating conditions: Ambient temperature: 10 °C – 40 °C
(50 °F – 104 °F)
Relative humidity: 30% – 85%
no condensation
Air pressure: 700 hPa – 1,060 hPa
Installation location: ≤ 3000 m above sea level
This treatment center is not suitable for
operation in areas subject to explosion
hazards.
Pollution degree: 2 acc. to IEC 60664-1
Load capacity of the patient
chair:
140 kg (308.6 lbs) or 185 kg (407.9 lbs),
depending on the model
The max. load capacity is indicated on a
sticker next to the rating plate of the
treatment center.
Tests/Approvals: See "Standards/Approvals“ [ → 23].
Year of manufacture:
(on the rating plate)
USB port: corresponds to USB 2.0 standard
IMPORTANT
Minimum requirements for PC
See document "Installation instructions and system requirements for PC
configuration," (REF 61 94 075) SIVISION digital.