Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.3Instruments and instrument hoses
båÖäáëÜ
2. Set all of the instruments to be purged to their maximum water flow
rate. The water regulators under the instrument holder must be
turned counterclockwise.
3. Put all instruments in place.
5.3.2.1
Purge function on the EasyPad
Setting purging time
The purging time of the removed instruments can be set between 20 and
120 seconds in the setup for the treatment center, see "Setting the
purging time for the purge function" [ 194].
Switching on the purging function
All instruments are in their holders.
1. Press the
Torque control/purge
key.
The
Purge
operating context is shown on the EasyPad display.
2. Press the favorite key
2 / Setup
.
Error message: Refill water (only for stand-alone water supply with
the Comfort water unit)
If the
Refill water
display appears on the EasyPad display after the purge
function has been started, there is not sufficient water in the disinfectant
tank of the water unit to purge the water lines. The purge function cannot
be started with insufficient water.
Mix distilled water with the DENTOSEPT P disinfectant at a ratio of
100:1 (1 liter of water, 10 ml DENTOSEPT P) and fill this into the
disinfectant tank of the water unit. For more information, please refer
to the section "Stand-alone water supply for the Comfort water unit".
When sufficient water has been refilled, the purge program
continues.
Purging water lines
A revolving element is shown on the EasyPad display next to the
Purge
display. Tumbler filling is automatically activated.
max.
max.
IMPORTANT
The water flow rate to the instruments is not checked by the treatment
center.
Make sure that the maximum water flow rate is set for all
instruments to be purged.
Purge
2
Setup