Sirona Dental Systems GmbH 4Operation
Operating Instructions INTEGO 4.5Dentist element
båÖäáëÜ
4.5.6 3-way standard and SPRAYVIT E syringe
The 3-way syringe is used to clean the treatment areas and blow them
dry. It supplies air and preheated water media. The heater for the
treatment water is in the water unit.
For the INTEGO, the water heater is optional.
4.5.6.1
Safety instructions
Multi-purpose syringe, nozzle falling out
Multifunctional syringe, water and air leakage
Water unit, water heater
IMPORTANT
Also note the SPRAYVIT operating instructions as well as the
infomation on care of the 3-way syringe Standard in the section
"Sterilizing the standard 3-way syringe" [ → 230].
WARNING
The nozzle of the Standard 3-way syringe must lock into the holder.
Otherwise, the jet could fall out during usage.
➢ Prior to usage, ensure that the jet is securely in place.
CAUTION
Leakage of air or water leak when the instrument or hose is replaced
If the Standard or SPRAYVIT E 3-way syringe is removed from the
instrument hose while the treatment center is in use, water and air will
escape from the hose coupling. When the instrument hose is unscrewed
from the treatment center, water and air also leaks out under the dentist
or assistant element.
Only the Comfort water unit with disinfection system has a valve that
prevents water leakage. Air can still escape.
➢ You should therefore switch off the treatment center before
changing an instrument or hose.
➢ Release any remaining air and water pressure by actuating a 3-way
syringe Standard or SPRAYVIT E before changing the hose or the
instrument.
IMPORTANT
Temperature monitoring of the water heater
If the electronic temperature monitor detects excessive heating of
treatment water (>42 °C / >107.6 F), the water heater is switched off and
an error message is given, see "Error messages" [ → 296].