5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.4Vacuum system Operating Instructions INTEGO
5.4.5 Empty the central suction sieve.
Depending on the options, there is a central suction sieve in the Compact
water unit or Comfort assistant element to retain solid materials (e.g.
amalgam). The suction sieve must be emptied when the suction force
deteriorates.
Amalgam residues, amalgam rotor an d sediment container
Wear gloves, risk of contamination
Empty suction siev e Compact or Turn wate r unit
Empty central suction sieve in the Compact or Ambidextrous water
unit
If the treatment center is equipped with the Compact assistant element,
the central suction sieve is in the swivel arm of the cuspidor. It is
accessible from the top.
1. Remove the cuspidor.
2. Remove the sieve cover and remove the sieve.
3. Collect amalgam from the suction sieve. Amalgam residues must be
disposed of separately.
4. Clean the sieve under running water and put it back into the water
unit.
CAUTION
Amalgam residues must not enter the public sewage system.
Amalgam is a mercury compound that is hazardous to water.
Do not dispose of amalgam residues in a sink.
Collect amalgam residues in a closed container with water. Dispose
of amalgam residues, e.g. when replacing the amalgam rotor by
filling the amalgam residues into the amalgam rotor or when
emptying the sediment container.
!
WARNING
Wear gloves when performing the following work.